
A dynamic model of polyethylene damage in dry total hip arthroplasties : wear and creep
- Author
- Ehsan Askari Renani (UGent) and Michael S. Andersen
- Organization
- Abstract
- The creep and wear of ultra-high-weight polyethylene hip prostheses under physiological conditions are studied in the present research work. Afully integrated contact-coupled dynamic model based upon multibody dynamics methodology is developed, allowing the evaluation of not only sliding distance, but also contact mechanics as well as cross-shear effects and both average pressure and in-service duration associated with the creep phenomenon. In vivo forces and motions of hip joint are used as input for the dynamic simulation, which result in more realistic contact point trajectory and contact pressure, and consequently wear and creep, compared to simplified inputs. The analysis also takes into account inertia forces due to hip motion, tribological properties of bearing bodies, and energy loss owing to contact-impact events. The principal molecular orientation (PMO) of the polyethylene cup is determined through an iterative algorithm and dynamic outcomes. Archard's wear law is also integrated into the multibody dynamics model for wear prediction in hip implants. Creep, besides wear, is attributed to polyethylene damage, which is investigated by implementing a creep model extracted from experimental data. The model is validated using clinical data and numerical results available from previously published studies. It is shown that creep plays a significant role in hip damage along with wear, both of which can be influenced by hip parameters, e.g., hip and clearance sizes. Moreover, the creep mechanism according to creep experiment is discussed, and contributing factors to the wear phenomenon are analyzed throughout this study.
- Keywords
- Multibody dynamics, Tribology, Total hip replacement, Contact mechanics
Downloads
-
(...).pdf
- full text (Published version)
- |
- UGent only
- |
- |
- 3.14 MB
Citation
Please use this url to cite or link to this publication: http://hdl.handle.net/1854/LU-8723111
- MLA
- Askari Renani, Ehsan, and Michael S. Andersen. “A Dynamic Model of Polyethylene Damage in Dry Total Hip Arthroplasties : Wear and Creep.” MULTIBODY SYSTEM DYNAMICS, vol. 45, no. 4, 2019, pp. 403–29, doi:10.1007/s11044-018-09652-2.
- APA
- Askari Renani, E., & Andersen, M. S. (2019). A dynamic model of polyethylene damage in dry total hip arthroplasties : wear and creep. MULTIBODY SYSTEM DYNAMICS, 45(4), 403–429. https://doi.org/10.1007/s11044-018-09652-2
- Chicago author-date
- Askari Renani, Ehsan, and Michael S. Andersen. 2019. “A Dynamic Model of Polyethylene Damage in Dry Total Hip Arthroplasties : Wear and Creep.” MULTIBODY SYSTEM DYNAMICS 45 (4): 403–29. https://doi.org/10.1007/s11044-018-09652-2.
- Chicago author-date (all authors)
- Askari Renani, Ehsan, and Michael S. Andersen. 2019. “A Dynamic Model of Polyethylene Damage in Dry Total Hip Arthroplasties : Wear and Creep.” MULTIBODY SYSTEM DYNAMICS 45 (4): 403–429. doi:10.1007/s11044-018-09652-2.
- Vancouver
- 1.Askari Renani E, Andersen MS. A dynamic model of polyethylene damage in dry total hip arthroplasties : wear and creep. MULTIBODY SYSTEM DYNAMICS. 2019;45(4):403–29.
- IEEE
- [1]E. Askari Renani and M. S. Andersen, “A dynamic model of polyethylene damage in dry total hip arthroplasties : wear and creep,” MULTIBODY SYSTEM DYNAMICS, vol. 45, no. 4, pp. 403–429, 2019.
@article{8723111, abstract = {{The creep and wear of ultra-high-weight polyethylene hip prostheses under physiological conditions are studied in the present research work. Afully integrated contact-coupled dynamic model based upon multibody dynamics methodology is developed, allowing the evaluation of not only sliding distance, but also contact mechanics as well as cross-shear effects and both average pressure and in-service duration associated with the creep phenomenon. In vivo forces and motions of hip joint are used as input for the dynamic simulation, which result in more realistic contact point trajectory and contact pressure, and consequently wear and creep, compared to simplified inputs. The analysis also takes into account inertia forces due to hip motion, tribological properties of bearing bodies, and energy loss owing to contact-impact events. The principal molecular orientation (PMO) of the polyethylene cup is determined through an iterative algorithm and dynamic outcomes. Archard's wear law is also integrated into the multibody dynamics model for wear prediction in hip implants. Creep, besides wear, is attributed to polyethylene damage, which is investigated by implementing a creep model extracted from experimental data. The model is validated using clinical data and numerical results available from previously published studies. It is shown that creep plays a significant role in hip damage along with wear, both of which can be influenced by hip parameters, e.g., hip and clearance sizes. Moreover, the creep mechanism according to creep experiment is discussed, and contributing factors to the wear phenomenon are analyzed throughout this study.}}, author = {{Askari Renani, Ehsan and Andersen, Michael S.}}, issn = {{1384-5640}}, journal = {{MULTIBODY SYSTEM DYNAMICS}}, keywords = {{Multibody dynamics,Tribology,Total hip replacement,Contact mechanics}}, language = {{eng}}, number = {{4}}, pages = {{403--429}}, title = {{A dynamic model of polyethylene damage in dry total hip arthroplasties : wear and creep}}, url = {{http://doi.org/10.1007/s11044-018-09652-2}}, volume = {{45}}, year = {{2019}}, }
- Altmetric
- View in Altmetric
- Web of Science
- Times cited: