Acute success and short-term follow-up of catheter ablation of isthmus-dependent atrial flutter: a comparison of 8 mm tip radiofrequency and cryothermy catheters

Author
Organization
Abstract
Objectives: To compare the acute success and short-term follow-up of ablation of atrial flutter using 8 mm tip radiofrequency (RF) and cryocatheters. Methods: Sixty-two patients with atrial flutter were randomized to RF or cryocatheter (cryo) ablation. Right atrial angiography was performed to assess the isthmus. End point was bidirectional isthmus block on multiple criteria. A pain score was used and the analgesics were recorded. Patients were followed for at least 3 months. Results: The acute success rate for RF was 83% vs 69% for cryo (NS). Procedure times were similar (mean 144 +/- 48 min for RF, vs 158 +/- 49 min for cryo). More applications were given with RF than with cryo (26 +/- 17 vs. 18 +/- 10, p +/- 0.05). Fluoroscopy time was longer with RF (29 +/- 15 vs. 19 +/- 12 min, p +/- 0.02). Peak CK, CK-MB and CK-MB mass were higher, also after 24 h in the cryo group. Troponin T did not differ. Repeated transient block during application (usually with cryoablation) seemed to predict failure. Cryothermy required significantly less analgesia (p +/- 0.01), and no use of long sheaths (p +/- 0.005). The isthmus tended to be longer in the failed procedures (p +/- 0.117). This was similar for both groups, as was the distribution of anatomic variations. Recurrences and complaints in the successful patients were similar for both groups, with a very low recurrence of atrial flutter after initial success. Conclusions: In this randomized study there was no statistical difference but a trend to less favorable outcome with 8 mm tip cryocatheters compared to RF catheters for atrial flutter ablation. Cryoablation was associated with less discomfort, fewer applications, shorter fluoroscopy times and similar procedure times. The recurrence rate was very low. Cryotherapy can be considered for atrial flutter ablation under certain circumstances especially when it has been used previously in the same patient, such as in an AF ablation.
Keywords
radiofrequency, cryothermy, atrial fibrillation, atrial flutter, catheter ablation, arrhythmia, CAVOTRICUSPID ISTHMUS, 8-MM-TIP CATHETERS, COOLED-TIP, CRYOABLATION, FIBRILLATION, EFFICACY, BLOCK, PREDICTORS, CONDUCTION, ANATOMY

Citation

Please use this url to cite or link to this publication:

MLA
Thornton, AS et al. “Acute Success and Short-term Follow-up of Catheter Ablation of Isthmus-dependent Atrial Flutter: a Comparison of 8 Mm Tip Radiofrequency and Cryothermy Catheters.” JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY 21.3 (2008): 241–248. Print.
APA
Thornton, A., Janse, P., Alings, M., Scholten, M., Mekel, J., Miltenburg, M., Jessurun, E., et al. (2008). Acute success and short-term follow-up of catheter ablation of isthmus-dependent atrial flutter: a comparison of 8 mm tip radiofrequency and cryothermy catheters. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 21(3), 241–248.
Chicago author-date
Thornton, AS, P Janse, M Alings, MF Scholten, JM Mekel, M Miltenburg, E Jessurun, and Luc Jordaens. 2008. “Acute Success and Short-term Follow-up of Catheter Ablation of Isthmus-dependent Atrial Flutter: a Comparison of 8 Mm Tip Radiofrequency and Cryothermy Catheters.” Journal of Interventional Cardiac Electrophysiology 21 (3): 241–248.
Chicago author-date (all authors)
Thornton, AS, P Janse, M Alings, MF Scholten, JM Mekel, M Miltenburg, E Jessurun, and Luc Jordaens. 2008. “Acute Success and Short-term Follow-up of Catheter Ablation of Isthmus-dependent Atrial Flutter: a Comparison of 8 Mm Tip Radiofrequency and Cryothermy Catheters.” Journal of Interventional Cardiac Electrophysiology 21 (3): 241–248.
Vancouver
1.
Thornton A, Janse P, Alings M, Scholten M, Mekel J, Miltenburg M, et al. Acute success and short-term follow-up of catheter ablation of isthmus-dependent atrial flutter: a comparison of 8 mm tip radiofrequency and cryothermy catheters. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY. 2008;21(3):241–8.
IEEE
[1]
A. Thornton et al., “Acute success and short-term follow-up of catheter ablation of isthmus-dependent atrial flutter: a comparison of 8 mm tip radiofrequency and cryothermy catheters,” JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, vol. 21, no. 3, pp. 241–248, 2008.
@article{8110297,
  abstract     = {Objectives: To compare the acute success and short-term follow-up of ablation of atrial flutter using 8 mm tip radiofrequency (RF) and cryocatheters. 
Methods: Sixty-two patients with atrial flutter were randomized to RF or cryocatheter (cryo) ablation. Right atrial angiography was performed to assess the isthmus. End point was bidirectional isthmus block on multiple criteria. A pain score was used and the analgesics were recorded. Patients were followed for at least 3 months. 
Results: The acute success rate for RF was 83% vs 69% for cryo (NS). Procedure times were similar (mean 144 +/- 48 min for RF, vs 158 +/- 49 min for cryo). More applications were given with RF than with cryo (26 +/- 17 vs. 18 +/- 10, p +/- 0.05). Fluoroscopy time was longer with RF (29 +/- 15 vs. 19 +/- 12 min, p +/- 0.02). Peak CK, CK-MB and CK-MB mass were higher, also after 24 h in the cryo group. Troponin T did not differ. Repeated transient block during application (usually with cryoablation) seemed to predict failure. Cryothermy required significantly less analgesia (p +/- 0.01), and no use of long sheaths (p +/- 0.005). 
The isthmus tended to be longer in the failed procedures (p +/- 0.117). This was similar for both groups, as was the distribution of anatomic variations. Recurrences and complaints in the successful patients were similar for both groups, with a very low recurrence of atrial flutter after initial success. 
Conclusions: In this randomized study there was no statistical difference but a trend to less favorable outcome with 8 mm tip cryocatheters compared to RF catheters for atrial flutter ablation. Cryoablation was associated with less discomfort, fewer applications, shorter fluoroscopy times and similar procedure times. The recurrence rate was very low. Cryotherapy can be considered for atrial flutter ablation under certain circumstances especially when it has been used previously in the same patient, such as in an AF ablation.},
  author       = {Thornton, AS and Janse, P and Alings, M and Scholten, MF and Mekel, JM and Miltenburg, M and Jessurun, E and Jordaens, Luc},
  issn         = {1383-875X},
  journal      = {JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY},
  keywords     = {radiofrequency,cryothermy,atrial fibrillation,atrial flutter,catheter ablation,arrhythmia,CAVOTRICUSPID ISTHMUS,8-MM-TIP CATHETERS,COOLED-TIP,CRYOABLATION,FIBRILLATION,EFFICACY,BLOCK,PREDICTORS,CONDUCTION,ANATOMY},
  language     = {eng},
  number       = {3},
  pages        = {241--248},
  title        = {Acute success and short-term follow-up of catheter ablation of isthmus-dependent atrial flutter: a comparison of 8 mm tip radiofrequency and cryothermy catheters},
  url          = {http://dx.doi.org/10.1007/s10840-008-9209-2},
  volume       = {21},
  year         = {2008},
}

Altmetric
View in Altmetric
Web of Science
Times cited: