Absorbed dose measurements in the build-up region of flattened versus unflattened megavoltage photon beams

Author
Organization
Abstract
This study evaluated absorbed dose measurements in the build-up region of conventional (FF) versus flattening filter-free (FFF) photon beams. The absorbed dose in the build-up region of static 6 and 10 MV FF and FFF beams was measured using radiochromic film and extrapolation chamber dosimetry for single beams with a variety of field sizes, shapes and positions relative to the central axis. Removing the flattening filter generally resulted in slightly higher relative build-up doses. No considerable impact on the depth of maximum dose was found.
Keywords
build-up dose, surface dose, Flattening filter-free, PARALLEL IONIZATION CHAMBERS, MEDICAL LINEAR-ACCELERATOR, FREE CLINICAL ACCELERATOR, MONTE-CARLO, DOSIMETRIC PROPERTIES, RADIOCHROMIC FILM, FILTER, QUALITY, IMRT

Downloads

  • (...).pdf
    • full text
    • |
    • UGent only
    • |
    • PDF
    • |
    • 673.95 KB

Citation

Please use this url to cite or link to this publication:

MLA
De Puysseleyr, Annemieke et al. “Absorbed Dose Measurements in the Build-up Region of Flattened Versus Unflattened Megavoltage Photon Beams.” ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK 26.2 (2016): 177–183. Print.
APA
De Puysseleyr, A., Lechner, W., De Neve, W., Georg, D., & De Wagter, C. (2016). Absorbed dose measurements in the build-up region of flattened versus unflattened megavoltage photon beams. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 26(2), 177–183.
Chicago author-date
De Puysseleyr, Annemieke, Wolfgang Lechner, Wilfried De Neve, Dietmar Georg, and Carlos De Wagter. 2016. “Absorbed Dose Measurements in the Build-up Region of Flattened Versus Unflattened Megavoltage Photon Beams.” Zeitschrift Fur Medizinische Physik 26 (2): 177–183.
Chicago author-date (all authors)
De Puysseleyr, Annemieke, Wolfgang Lechner, Wilfried De Neve, Dietmar Georg, and Carlos De Wagter. 2016. “Absorbed Dose Measurements in the Build-up Region of Flattened Versus Unflattened Megavoltage Photon Beams.” Zeitschrift Fur Medizinische Physik 26 (2): 177–183.
Vancouver
1.
De Puysseleyr A, Lechner W, De Neve W, Georg D, De Wagter C. Absorbed dose measurements in the build-up region of flattened versus unflattened megavoltage photon beams. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK. 2016;26(2):177–83.
IEEE
[1]
A. De Puysseleyr, W. Lechner, W. De Neve, D. Georg, and C. De Wagter, “Absorbed dose measurements in the build-up region of flattened versus unflattened megavoltage photon beams,” ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, vol. 26, no. 2, pp. 177–183, 2016.
@article{8056140,
  abstract     = {This study evaluated absorbed dose measurements in the build-up region of conventional (FF) versus flattening filter-free (FFF) photon beams. The absorbed dose in the build-up region of static 6 and 10 MV FF and FFF beams was measured using radiochromic film and extrapolation chamber dosimetry for single beams with a variety of field sizes, shapes and positions relative to the central axis. Removing the flattening filter generally resulted in slightly higher relative build-up doses. No considerable impact on the depth of maximum dose was found.},
  author       = {De Puysseleyr, Annemieke and Lechner, Wolfgang and De Neve, Wilfried and Georg, Dietmar and De Wagter, Carlos},
  issn         = {0939-3889},
  journal      = {ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK},
  keywords     = {build-up dose,surface dose,Flattening filter-free,PARALLEL IONIZATION CHAMBERS,MEDICAL LINEAR-ACCELERATOR,FREE CLINICAL ACCELERATOR,MONTE-CARLO,DOSIMETRIC PROPERTIES,RADIOCHROMIC FILM,FILTER,QUALITY,IMRT},
  language     = {eng},
  number       = {2},
  pages        = {177--183},
  title        = {Absorbed dose measurements in the build-up region of flattened versus unflattened megavoltage photon beams},
  url          = {http://dx.doi.org/10.1016/j.zemedi.2016.02.005},
  volume       = {26},
  year         = {2016},
}

Altmetric
View in Altmetric
Web of Science
Times cited: