Monte Carlo evaluation of the AAA treatment planning algorithm in a heterogeneous multilayer phantom and IMRT clinical treatments for an Elekta SL25 linear accelerator

E. Sterpin, M. Tomsej, B. De Smedt, N. Reynaert and S. Vynckier (2007) MEDICAL PHYSICS. 34(5). p.1665-1677
abstract
The Anisotropic Analytical Algorithm (AAA) is a new pencil beam convolution/superposition algorithm proposed by Varian for photon dose calculations. The configuration of AAA depends on linear accelerator design and specifications. The purpose of this study was to investigate the accuracy of AAA for an Elekta SL25 linear accelerator for small fields and intensity modulated radiation therapy (IMRT) treatments in inhomogeneous media. The accuracy of AAA was evaluated in two studies. First, AAA was compared both with Monte Carlo (MC) and the measurements in an inhomogeneous phantom simulating lung equivalent tissues and bone ribs. The algorithm was tested under lateral electronic disequilibrium conditions, using small fields (2 x 2 cm(2)). Good agreement was generally achieved for depth dose and profiles, with deviations generally below 3% in lung inhomogeneities and below 5% at interfaces. However, the effects of attenuation and scattering close to the bone ribs were not fully taken into account by AAA, and small inhomogeneities may lead to planning errors. Second, AAA and MC were compared for IMRT plans in clinical conditions, i.e., dose calculations in a computed tomography scan of a patient. One ethmoid tumor, one orophaxynx and two lung tumors are presented in this paper. Small differences were found between the dose volume histograms. For instance, a 1.7% difference for the mean planning target volume dose was obtained for the ethmoid case. Since better agreement was achieved for the same plans but in homogeneous conditions, these differences must be attributed to the handling of inhomogeneities by AAA. Therefore, inherent assumptions of the algorithm, principally the assumption of independent depth and lateral directions in the scaling of the kernels, were slightly influencing AAA's validity in inhomogeneities. However, AAA showed a good accuracy overall and a great ability to handle small fields in inhomogeneous media compared to other pencil beam convolution algorithms.
Please use this url to cite or link to this publication:
author
organization
year
type
journalArticle (original)
publication status
published
keyword
ELECTRON-TRANSPORT, COLLAPSED CONE CONVOLUTION, QUALITY-ASSURANCE, RADIOGRAPHIC FILM, KODAK EDR2, BEAM, VERIFICATION, DOSIMETRY, PHOTON DOSE CALCULATION, MODULATED RADIATION-THERAPY
journal title
MEDICAL PHYSICS
Med. Phys.
volume
34
issue
5
pages
1665 - 1677
Web of Science type
Article
Web of Science id
000246629800016
JCR category
RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
JCR impact factor
3.198 (2007)
JCR rank
16/86 (2007)
JCR quartile
1 (2007)
ISSN
0094-2405
DOI
10.1118/1.2727314
language
English
UGent publication?
yes
classification
A1
id
743074
handle
http://hdl.handle.net/1854/LU-743074
date created
2009-09-09 08:59:39
date last changed
2009-10-19 14:53:28
@article{743074,
  abstract     = {The Anisotropic Analytical Algorithm (AAA) is a new pencil beam convolution/superposition algorithm proposed by Varian for photon dose calculations. The configuration of AAA depends on linear accelerator design and specifications. The purpose of this study was to investigate the accuracy of AAA for an Elekta SL25 linear accelerator for small fields and intensity modulated radiation therapy (IMRT) treatments in inhomogeneous media. The accuracy of AAA was evaluated in two studies. First, AAA was compared both with Monte Carlo (MC) and the measurements in an inhomogeneous phantom simulating lung equivalent tissues and bone ribs. The algorithm was tested under lateral electronic disequilibrium conditions, using small fields (2 x 2 cm(2)). Good agreement was generally achieved for depth dose and profiles, with deviations generally below 3\% in lung inhomogeneities and below 5\% at interfaces. However, the effects of attenuation and scattering close to the bone ribs were not fully taken into account by AAA, and small inhomogeneities may lead to planning errors. Second, AAA and MC were compared for IMRT plans in clinical conditions, i.e., dose calculations in a computed tomography scan of a patient. One ethmoid tumor, one orophaxynx and two lung tumors are presented in this paper. Small differences were found between the dose volume histograms. For instance, a 1.7\% difference for the mean planning target volume dose was obtained for the ethmoid case. Since better agreement was achieved for the same plans but in homogeneous conditions, these differences must be attributed to the handling of inhomogeneities by AAA. Therefore, inherent assumptions of the algorithm, principally the assumption of independent depth and lateral directions in the scaling of the kernels, were slightly influencing AAA's validity in inhomogeneities. However, AAA showed a good accuracy overall and a great ability to handle small fields in inhomogeneous media compared to other pencil beam convolution algorithms.},
  author       = {Sterpin, E. and Tomsej, M. and De Smedt, B. and Reynaert, N. and Vynckier, S.},
  issn         = {0094-2405},
  journal      = {MEDICAL PHYSICS},
  keyword      = {ELECTRON-TRANSPORT,COLLAPSED CONE CONVOLUTION,QUALITY-ASSURANCE,RADIOGRAPHIC FILM,KODAK EDR2,BEAM,VERIFICATION,DOSIMETRY,PHOTON DOSE CALCULATION,MODULATED RADIATION-THERAPY},
  language     = {eng},
  number       = {5},
  pages        = {1665--1677},
  title        = {Monte Carlo evaluation of the AAA treatment planning algorithm in a heterogeneous multilayer phantom and IMRT clinical treatments for an Elekta SL25 linear accelerator},
  url          = {http://dx.doi.org/10.1118/1.2727314},
  volume       = {34},
  year         = {2007},
}

Chicago
Sterpin, E., M. Tomsej, B. De Smedt, N. Reynaert, and S. Vynckier. 2007. “Monte Carlo Evaluation of the AAA Treatment Planning Algorithm in a Heterogeneous Multilayer Phantom and IMRT Clinical Treatments for an Elekta SL25 Linear Accelerator.” Medical Physics 34 (5): 1665–1677.
APA
Sterpin, E., Tomsej, M., De Smedt, B., Reynaert, N., & Vynckier, S. (2007). Monte Carlo evaluation of the AAA treatment planning algorithm in a heterogeneous multilayer phantom and IMRT clinical treatments for an Elekta SL25 linear accelerator. MEDICAL PHYSICS, 34(5), 1665–1677.
Vancouver
1.
Sterpin E, Tomsej M, De Smedt B, Reynaert N, Vynckier S. Monte Carlo evaluation of the AAA treatment planning algorithm in a heterogeneous multilayer phantom and IMRT clinical treatments for an Elekta SL25 linear accelerator. MEDICAL PHYSICS. 2007;34(5):1665–77.
MLA
Sterpin, E., M. Tomsej, B. De Smedt, et al. “Monte Carlo Evaluation of the AAA Treatment Planning Algorithm in a Heterogeneous Multilayer Phantom and IMRT Clinical Treatments for an Elekta SL25 Linear Accelerator.” MEDICAL PHYSICS 34.5 (2007): 1665–1677. Print.