Advanced search

Comparison of 3D SPECT imaging with a rotating slat collimator and a parallel hole collimator

Roel Van Holen (UGent) , Stefaan Vandenberghe (UGent) , Steven Staelens (UGent) and Ignace Lemahieu (UGent)
Author
Organization
Abstract
Besides parallel and converging hole collimators which are frequently used in nuclear medicine, slat collimators can be used for Single Photon Emission Computed Tomography (SPECT). The higher photon collection efficiency, inherent to the geometry of rotating slat collimators results in much lower noise in the data. However, reconstruction of the plane integral data measured by a slat collimator is much more sensitive to noise accumulation compared to traditional SPECT reconstruction from line integrals. It is not a straightforward question whether the initial gain in efficiency will compensate for the larger noise accumulation during reconstruction. Therefore, a comparison of the performance of parallel hole and rotating slat collimation is needed. This study compares SPECT with rotating slat and parallel hole collimation in combination with MLEM reconstruction with accurate system modeling and correction for scatter and attenuation. A contrast-to-noise study revealed an improvement of a factor 3 to 4 for hot lesions and more than a factor of 4 for cold lesion. Rotating slat collimators are thus a valuable alternative for parallel hole collimators.
Keywords
slat collimator, SPECT, COMPUTED-TOMOGRAPHY, RECONSTRUCTION, SYSTEM, ATTENUATION, DETECTOR

Citation

Please use this url to cite or link to this publication:

Chicago
Van Holen, Roel, Stefaan Vandenberghe, Steven Staelens, and Ignace Lemahieu. 2009. “Comparison of 3D SPECT Imaging with a Rotating Slat Collimator and a Parallel Hole Collimator.” In Ieee Nuclear Science Symposium - Conference Record, 3865–3870. New York, NY, USA: IEEE.
APA
Van Holen, R., Vandenberghe, S., Staelens, S., & Lemahieu, I. (2009). Comparison of 3D SPECT imaging with a rotating slat collimator and a parallel hole collimator. IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD (pp. 3865–3870). Presented at the IEEE Nuclear Science Symposium/Medical Imaging Conference, New York, NY, USA: IEEE.
Vancouver
1.
Van Holen R, Vandenberghe S, Staelens S, Lemahieu I. Comparison of 3D SPECT imaging with a rotating slat collimator and a parallel hole collimator. IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD. New York, NY, USA: IEEE; 2009. p. 3865–70.
MLA
Van Holen, Roel, Stefaan Vandenberghe, Steven Staelens, et al. “Comparison of 3D SPECT Imaging with a Rotating Slat Collimator and a Parallel Hole Collimator.” Ieee Nuclear Science Symposium - Conference Record. New York, NY, USA: IEEE, 2009. 3865–3870. Print.
@inproceedings{677829,
  abstract     = {Besides parallel and converging hole collimators which are frequently used in nuclear medicine, slat collimators can be used for Single Photon Emission Computed Tomography (SPECT). The higher photon collection efficiency, inherent to the geometry of rotating slat collimators results in much lower noise in the data. However, reconstruction of the plane integral data measured by a slat collimator is much more sensitive to noise accumulation compared to traditional SPECT reconstruction from line integrals. It is not a straightforward question whether the initial gain in efficiency will compensate for the larger noise accumulation during reconstruction. Therefore, a comparison of the performance of parallel hole and rotating slat collimation is needed. This study compares SPECT with rotating slat and parallel hole collimation in combination with MLEM reconstruction with accurate system modeling and correction for scatter and attenuation. A contrast-to-noise study revealed an improvement of a factor 3 to 4 for hot lesions and more than a factor of 4 for cold lesion. Rotating slat collimators are thus a valuable alternative for parallel hole collimators.},
  author       = {Van Holen, Roel and Vandenberghe, Stefaan and Staelens, Steven and Lemahieu, Ignace},
  booktitle    = {IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD},
  isbn         = {978-1-4244-2714-7},
  issn         = {1082-3654},
  language     = {eng},
  location     = {Dresden, Germany},
  pages        = {3865--3870},
  publisher    = {IEEE},
  title        = {Comparison of 3D SPECT imaging with a rotating slat collimator and a parallel hole collimator},
  url          = {http://dx.doi.org/10.1109/NSSMIC.2008.4774366},
  year         = {2009},
}