Advanced search
1 file | 2.57 MB Add to list

Soft, comfortable polymer dry electrodes for high quality ECG and EEG recording

(2014) SENSORS. 14(12). p.23758-23780
Author
Organization
Abstract
Conventional gel electrodes are widely used for biopotential measurements, despite important drawbacks such as skin irritation, long set-up time and uncomfortable removal. Recently introduced dry electrodes with rigid metal pins overcome most of these problems; however, their rigidity causes discomfort and pain. This paper presents dry electrodes offering high user comfort, since they are fabricated from EPDM rubber containing various additives for optimum conductivity, flexibility and ease of fabrication. The electrode impedance is measured on phantoms and human skin. After optimization of the polymer composition, the skin-electrode impedance is only similar to 10 times larger than that of gel electrodes. Therefore, these electrodes are directly capable of recording strong biopotential signals such as ECG while for low-amplitude signals such as EEG, the electrodes need to be coupled with an active circuit. EEG recordings using active polymer electrodes connected to a clinical EEG system show very promising results: alpha waves can be clearly observed when subjects close their eyes, and correlation and coherence analyses reveal high similarity between dry and gel electrode signals. Moreover, all subjects reported that our polymer electrodes did not cause discomfort. Hence, the polymer-based dry electrodes are promising alternatives to either rigid dry electrodes or conventional gel electrodes.
Keywords
material optimization, flexible polymer dry electrode, impedance, high quality biopotential recordings, ECG, EEG, high user comfort, conductive polymer, IMPEDANCE, ELECTROCARDIOGRAPHY, SYSTEMS, SENSOR

Downloads

  • Chen-2015.pdf
    • full text
    • |
    • open access
    • |
    • PDF
    • |
    • 2.57 MB

Citation

Please use this url to cite or link to this publication:

MLA
Chen, Yun-Hsuan et al. “Soft, Comfortable Polymer Dry Electrodes for High Quality ECG and EEG Recording.” SENSORS 14.12 (2014): 23758–23780. Print.
APA
Chen, Y.-H., Op de Beeck, M., Vanderheyden, L., Carrette, E., Mihajlović, V., Vanstreels, K., Grundlehner, B., et al. (2014). Soft, comfortable polymer dry electrodes for high quality ECG and EEG recording. SENSORS, 14(12), 23758–23780.
Chicago author-date
Chen, Yun-Hsuan, Maaike Op de Beeck, Luc Vanderheyden, Evelien Carrette, Vojkan Mihajlović, Kris Vanstreels, Bernard Grundlehner, Stefanie Gadeyne, Paul Boon, and Chris Van Hoof. 2014. “Soft, Comfortable Polymer Dry Electrodes for High Quality ECG and EEG Recording.” Sensors 14 (12): 23758–23780.
Chicago author-date (all authors)
Chen, Yun-Hsuan, Maaike Op de Beeck, Luc Vanderheyden, Evelien Carrette, Vojkan Mihajlović, Kris Vanstreels, Bernard Grundlehner, Stefanie Gadeyne, Paul Boon, and Chris Van Hoof. 2014. “Soft, Comfortable Polymer Dry Electrodes for High Quality ECG and EEG Recording.” Sensors 14 (12): 23758–23780.
Vancouver
1.
Chen Y-H, Op de Beeck M, Vanderheyden L, Carrette E, Mihajlović V, Vanstreels K, et al. Soft, comfortable polymer dry electrodes for high quality ECG and EEG recording. SENSORS. 2014;14(12):23758–80.
IEEE
[1]
Y.-H. Chen et al., “Soft, comfortable polymer dry electrodes for high quality ECG and EEG recording,” SENSORS, vol. 14, no. 12, pp. 23758–23780, 2014.
@article{5908184,
  abstract     = {Conventional gel electrodes are widely used for biopotential measurements, despite important drawbacks such as skin irritation, long set-up time and uncomfortable removal. Recently introduced dry electrodes with rigid metal pins overcome most of these problems; however, their rigidity causes discomfort and pain. This paper presents dry electrodes offering high user comfort, since they are fabricated from EPDM rubber containing various additives for optimum conductivity, flexibility and ease of fabrication. The electrode impedance is measured on phantoms and human skin. After optimization of the polymer composition, the skin-electrode impedance is only similar to 10 times larger than that of gel electrodes. Therefore, these electrodes are directly capable of recording strong biopotential signals such as ECG while for low-amplitude signals such as EEG, the electrodes need to be coupled with an active circuit. EEG recordings using active polymer electrodes connected to a clinical EEG system show very promising results: alpha waves can be clearly observed when subjects close their eyes, and correlation and coherence analyses reveal high similarity between dry and gel electrode signals. Moreover, all subjects reported that our polymer electrodes did not cause discomfort. Hence, the polymer-based dry electrodes are promising alternatives to either rigid dry electrodes or conventional gel electrodes.},
  author       = {Chen, Yun-Hsuan and Op de Beeck, Maaike and Vanderheyden, Luc and Carrette, Evelien and Mihajlović, Vojkan and Vanstreels, Kris and Grundlehner, Bernard and Gadeyne, Stefanie and Boon, Paul and Van Hoof, Chris},
  issn         = {1424-8220},
  journal      = {SENSORS},
  keywords     = {material optimization,flexible polymer dry electrode,impedance,high quality biopotential recordings,ECG,EEG,high user comfort,conductive polymer,IMPEDANCE,ELECTROCARDIOGRAPHY,SYSTEMS,SENSOR},
  language     = {eng},
  number       = {12},
  pages        = {23758--23780},
  title        = {Soft, comfortable polymer dry electrodes for high quality ECG and EEG recording},
  url          = {http://dx.doi.org/10.3390/s141223758},
  volume       = {14},
  year         = {2014},
}

Altmetric
View in Altmetric
Web of Science
Times cited: