Impact of the digitalisation of mammography on performance parameters and breast dose in the Flemish Breast Cancer Screening Programme

(2014) EUROPEAN RADIOLOGY. 24(8). p.1808-1819
Author
Organization
Abstract
To investigate the impact of digitalisation on performance parameters and breast dose of the Flemish Breast Cancer Screening Programme. Both computed (CR) and direct radiography (DR) are compared with screen-film mammography (SFM). Data from 975,673 mammographic examinations were collected from units which underwent digitalisation from SFM to CR (41 units) or DR (72 units) in the period 2005-2011. Performance indicators were obtained by consulting the Screening Programme database. Phantom and patient dosimetry data were acquired from the physical technical quality assurance of the programme. Digitalisation induced no significant change in cancer detection rate (CDR), percentage of ductal carcinomas in situ and percentage of breast cancers smaller than 1 cm. A decrease in false-positive results and third readings was observed, which was a time-related observation. After digitalisation, positive predictive value (PPV) increased and recall rates decreased. Compared with SFM, an increase of 30 % in mean glandular dose (MGD) was found for CR, while a similar change in the opposite direction was found for DR. No major differences in performance parameters after digitalisation were found. Transition of SFM to CR resulted in a higher MGD and associated lower detection-over-induction ratio (DIR), while the change to DR induced an improvement of DIR. * Performance parameters showed no major differences after digitalisation to CR or DR. * Transition from SFM to CR results in a higher mean glandular dose. * Transition from SFM to DR results in a lower mean glandular dose.
Keywords
FILM MAMMOGRAPHY, Glandular dose, COMPUTED RADIOGRAPHY, X-RAYS, CLASSIFICATION, DETECTOR, BENEFIT, COHORTS, TRIAL, RISK, Performance parameters, Digital mammography, Screen-Film mammography, Breast cancer screening

Downloads

  • (...).pdf
    • full text
    • |
    • UGent only
    • |
    • PDF
    • |
    • 982.10 KB

Citation

Please use this url to cite or link to this publication:

MLA
Timmermans, Lore, et al. “Impact of the Digitalisation of Mammography on Performance Parameters and Breast Dose in the Flemish Breast Cancer Screening Programme.” EUROPEAN RADIOLOGY, vol. 24, no. 8, 2014, pp. 1808–19, doi:10.1007/s00330-014-3169-y.
APA
Timmermans, L., De Hauwere, A., Bacher, K., Bosmans, H., Lemmens, K., Bleyen, L., … Thierens, H. (2014). Impact of the digitalisation of mammography on performance parameters and breast dose in the Flemish Breast Cancer Screening Programme. EUROPEAN RADIOLOGY, 24(8), 1808–1819. https://doi.org/10.1007/s00330-014-3169-y
Chicago author-date
Timmermans, Lore, An De Hauwere, Klaus Bacher, Hilde Bosmans, Kim Lemmens, Luc Bleyen, Erik Van Limbergen, et al. 2014. “Impact of the Digitalisation of Mammography on Performance Parameters and Breast Dose in the Flemish Breast Cancer Screening Programme.” EUROPEAN RADIOLOGY 24 (8): 1808–19. https://doi.org/10.1007/s00330-014-3169-y.
Chicago author-date (all authors)
Timmermans, Lore, An De Hauwere, Klaus Bacher, Hilde Bosmans, Kim Lemmens, Luc Bleyen, Erik Van Limbergen, Patrick Martens, Andre Van Steen, Margarete Mortier, Koen Van Herck, and Hubert Thierens. 2014. “Impact of the Digitalisation of Mammography on Performance Parameters and Breast Dose in the Flemish Breast Cancer Screening Programme.” EUROPEAN RADIOLOGY 24 (8): 1808–1819. doi:10.1007/s00330-014-3169-y.
Vancouver
1.
Timmermans L, De Hauwere A, Bacher K, Bosmans H, Lemmens K, Bleyen L, et al. Impact of the digitalisation of mammography on performance parameters and breast dose in the Flemish Breast Cancer Screening Programme. EUROPEAN RADIOLOGY. 2014;24(8):1808–19.
IEEE
[1]
L. Timmermans et al., “Impact of the digitalisation of mammography on performance parameters and breast dose in the Flemish Breast Cancer Screening Programme,” EUROPEAN RADIOLOGY, vol. 24, no. 8, pp. 1808–1819, 2014.
@article{5683396,
  abstract     = {{To investigate the impact of digitalisation on performance parameters and breast dose of the Flemish Breast Cancer Screening Programme. Both computed (CR) and direct radiography (DR) are compared with screen-film mammography (SFM).
Data from 975,673 mammographic examinations were collected from units which underwent digitalisation from SFM to CR (41 units) or DR (72 units) in the period 2005-2011. Performance indicators were obtained by consulting the Screening Programme database. Phantom and patient dosimetry data were acquired from the physical technical quality assurance of the programme.
Digitalisation induced no significant change in cancer detection rate (CDR), percentage of ductal carcinomas in situ and percentage of breast cancers smaller than 1 cm. A decrease in false-positive results and third readings was observed, which was a time-related observation. After digitalisation, positive predictive value (PPV) increased and recall rates decreased. Compared with SFM, an increase of 30 % in mean glandular dose (MGD) was found for CR, while a similar change in the opposite direction was found for DR.
No major differences in performance parameters after digitalisation were found. Transition of SFM to CR resulted in a higher MGD and associated lower detection-over-induction ratio (DIR), while the change to DR induced an improvement of DIR.
* Performance parameters showed no major differences after digitalisation to CR or DR.
* Transition from SFM to CR results in a higher mean glandular dose. 
* Transition from SFM to DR results in a lower mean glandular dose.}},
  author       = {{Timmermans, Lore and De Hauwere, An and Bacher, Klaus and Bosmans, Hilde and Lemmens, Kim and Bleyen, Luc and Van Limbergen, Erik and Martens, Patrick and Van Steen, Andre and Mortier, Margarete and Van Herck, Koen and Thierens, Hubert}},
  issn         = {{0938-7994}},
  journal      = {{EUROPEAN RADIOLOGY}},
  keywords     = {{FILM MAMMOGRAPHY,Glandular dose,COMPUTED RADIOGRAPHY,X-RAYS,CLASSIFICATION,DETECTOR,BENEFIT,COHORTS,TRIAL,RISK,Performance parameters,Digital mammography,Screen-Film mammography,Breast cancer screening}},
  language     = {{eng}},
  number       = {{8}},
  pages        = {{1808--1819}},
  title        = {{Impact of the digitalisation of mammography on performance parameters and breast dose in the Flemish Breast Cancer Screening Programme}},
  url          = {{http://doi.org/10.1007/s00330-014-3169-y}},
  volume       = {{24}},
  year         = {{2014}},
}

Altmetric
View in Altmetric
Web of Science
Times cited: