Comparison study of RPC and crystal based PET systems for hadron therapy monitoring

Author
Organization
Abstract
PET is the only clinically proven method for hadron-therapy monitoring. This method uses imaging of the positron emitting isotopes produced during patient irradiation in hadron-therapy. To maximize the count rates in-beam PET is the preferred option, which requires partial ring configuration. Usage of partial ring PET system causes image artifact and low quality image due to incomplete data. The Time-of-flight (TOF) information provides better image quality for partial ring PET systems. Therefore, PET system with very good time and spatial resolution is necessary for hadron-therapy monitoring. The Resistive-Plate-Chambers (RPCs) providing very good time and spatial resolution might be used for the quality control of hadron-therapy. The objective of this study is the comparison of the performance of two TOF-PET systems, one multi-gap RPC (mRPC) system currently under investigation by the TERA group and a commercially available scintillator-based PET system (Philips GEMINI TF) in order to analyze their feasibility of being applied to monitor cancer treatment with ion- and proton-beams. Various Monte-Carlo simulations were performed using GATE to evaluate the performance of the both systems following NEMA protocols. Furthermore, a set of sources emulating the annihilation photons from β+-emitters produced after proton irradiation on a PMMA target were employed to study the capability to reproduce the proton beam path and the distal edge. Finally, we calculated the activity distribution produced on the basis of one field of the treatment plan for one patient data. The results obtained indicate that mRPC system is not suited for hadron therapy monitoring due to its low sensitivity. On the other side, the Philips GEMINI TF does not provide convincing results either, thus other configurations are still under investigation.
Keywords
Hadron therapy, Resistive Plate Chamber, Gemini TF, RPC, TOF-PET

Downloads

  • (...).pdf
    • full text
    • |
    • UGent only
    • |
    • PDF
    • |
    • 2.81 MB

Citation

Please use this url to cite or link to this publication:

Chicago
Diblen, Faruk, Heide Rohling, Irene Torres-Espallardo, Paola Solevi, John Gillam, Samuel Espana Palomares, Stefaan Vandenberghe, David Watts, Fine Fiedler, and Magdalena Refecas. 2012. “Comparison Study of RPC and Crystal Based PET Systems for Hadron Therapy Monitoring.” In IEEE Nuclear Science Symposium Conference Record, ed. B Yu, 2212–2217. New York, NY, USA: IEEE.
APA
Diblen, F., Rohling, H., Torres-Espallardo, I., Solevi, P., Gillam, J., Espana Palomares, S., Vandenberghe, S., et al. (2012). Comparison study of RPC and crystal based PET systems for hadron therapy monitoring. In B Yu (Ed.), IEEE Nuclear Science Symposium Conference Record (pp. 2212–2217). Presented at the 2012 IEEE Nuclear Science symposium and Medical Imaging Conference record (NSS/MIC 2012) ; Workshop on Room-Temperature Semiconductor X-Ray and Gamma-Ray Detectors, New York, NY, USA: IEEE.
Vancouver
1.
Diblen F, Rohling H, Torres-Espallardo I, Solevi P, Gillam J, Espana Palomares S, et al. Comparison study of RPC and crystal based PET systems for hadron therapy monitoring. In: Yu B, editor. IEEE Nuclear Science Symposium Conference Record. New York, NY, USA: IEEE; 2012. p. 2212–7.
MLA
Diblen, Faruk, Heide Rohling, Irene Torres-Espallardo, et al. “Comparison Study of RPC and Crystal Based PET Systems for Hadron Therapy Monitoring.” IEEE Nuclear Science Symposium Conference Record. Ed. B Yu. New York, NY, USA: IEEE, 2012. 2212–2217. Print.
@inproceedings{3076241,
  abstract     = {PET is the only clinically proven method for hadron-therapy monitoring. This method uses imaging of the positron emitting isotopes produced during patient irradiation in hadron-therapy. To maximize the count rates in-beam PET is the preferred option, which requires partial ring configuration. Usage of partial ring PET system causes image artifact and low quality image due to incomplete data. The Time-of-flight (TOF) information provides better image quality for partial ring PET systems. Therefore, PET system with very good time and spatial resolution is necessary for hadron-therapy monitoring. The Resistive-Plate-Chambers (RPCs) providing very good time and spatial resolution might be used for the quality control of hadron-therapy. The objective of this study is the comparison of the performance of two TOF-PET systems, one multi-gap RPC (mRPC) system currently under investigation by the TERA group and a commercially available scintillator-based PET system (Philips GEMINI TF) in order to analyze their feasibility of being applied to monitor cancer treatment with ion- and proton-beams. Various Monte-Carlo simulations were performed using GATE to evaluate the performance of the both systems following NEMA protocols. Furthermore, a set of sources emulating the annihilation photons from \ensuremath{\beta}+-emitters produced after proton irradiation on a PMMA target were employed to study the capability to reproduce the proton beam path and the distal edge. Finally, we calculated the activity distribution produced on the basis of one field of the treatment plan for one patient data. The results obtained indicate that mRPC system is not suited for hadron therapy monitoring due to its low sensitivity. On the other side, the Philips GEMINI TF does not provide convincing results either, thus other configurations are still under investigation.},
  author       = {Diblen, Faruk and Rohling, Heide and Torres-Espallardo, Irene and Solevi, Paola and Gillam, John and Espana Palomares, Samuel and Vandenberghe, Stefaan and Watts, David and Fiedler, Fine and Refecas, Magdalena},
  booktitle    = {IEEE Nuclear Science Symposium Conference Record},
  editor       = {Yu, B},
  isbn         = {9781467320283},
  issn         = {1082-3654},
  language     = {eng},
  location     = {Anaheim, CA, USA},
  pages        = {2212--2217},
  publisher    = {IEEE},
  title        = {Comparison study of RPC and crystal based PET systems for hadron therapy monitoring},
  url          = {http://dx.doi.org/10.1109/NSSMIC.2012.6551504},
  year         = {2012},
}

Altmetric
View in Altmetric
Web of Science
Times cited: