Atrial lead malfunction presenting as new onset pacemaker-mediated tachycardia

S Serge Barold and Roland Stroobandt UGent (2012) EUROPACE. 14(7). p.1060-1061
abstract
This report describes the de novo occurrence of pacemaker-mediated tachycardia (PMT) in a patient with a dual-chamber implantable cardioverter-defibrillator and stable retrograde ventriculoatrial conduction time. The same rate-adaptive post-ventricular atrial refractory period (PVARP) duration had previously prevented PMT. Oversensing of atrial false signals from a defective lead shortened the PVARP with consequent sensing of retrograde conduction.
Please use this url to cite or link to this publication:
author
organization
year
type
journalArticle (original)
publication status
published
subject
keyword
Cardiac pacemaker, Malfunction, Pacemaker-mediated tachycardia (PMT), Atrial lead
journal title
EUROPACE
Europace
volume
14
issue
7
pages
1060 - 1061
Web of Science type
Article
Web of Science id
000305621000026
JCR category
CARDIAC & CARDIOVASCULAR SYSTEMS
JCR impact factor
2.765 (2012)
JCR rank
45/120 (2012)
JCR quartile
2 (2012)
ISSN
1099-5129
DOI
10.1093/europace/eur405
language
English
UGent publication?
yes
classification
A1
copyright statement
I have transferred the copyright for this publication to the publisher
id
1976426
handle
http://hdl.handle.net/1854/LU-1976426
date created
2012-01-01 15:35:26
date last changed
2017-05-04 12:12:11
@article{1976426,
  abstract     = {This report describes the de novo occurrence of pacemaker-mediated tachycardia (PMT) in a patient with a dual-chamber implantable cardioverter-defibrillator and stable retrograde ventriculoatrial conduction time. The same rate-adaptive post-ventricular atrial refractory period (PVARP) duration had previously prevented PMT. Oversensing of atrial false signals from a defective lead shortened the PVARP with consequent sensing of retrograde conduction.},
  author       = {Barold, S Serge and Stroobandt, Roland},
  issn         = {1099-5129},
  journal      = {EUROPACE},
  keyword      = {Cardiac pacemaker,Malfunction,Pacemaker-mediated tachycardia (PMT),Atrial lead},
  language     = {eng},
  number       = {7},
  pages        = {1060--1061},
  title        = {Atrial lead malfunction presenting as new onset pacemaker-mediated tachycardia},
  url          = {http://dx.doi.org/10.1093/europace/eur405},
  volume       = {14},
  year         = {2012},
}

Chicago
Barold, S Serge, and Roland Stroobandt. 2012. “Atrial Lead Malfunction Presenting as New Onset Pacemaker-mediated Tachycardia.” Europace 14 (7): 1060–1061.
APA
Barold, S. S., & Stroobandt, R. (2012). Atrial lead malfunction presenting as new onset pacemaker-mediated tachycardia. EUROPACE, 14(7), 1060–1061.
Vancouver
1.
Barold SS, Stroobandt R. Atrial lead malfunction presenting as new onset pacemaker-mediated tachycardia. EUROPACE. 2012;14(7):1060–1.
MLA
Barold, S Serge, and Roland Stroobandt. “Atrial Lead Malfunction Presenting as New Onset Pacemaker-mediated Tachycardia.” EUROPACE 14.7 (2012): 1060–1061. Print.