Silent lead malfunction detected only during defibrillator replacement

Roland Stroobandt UGent, Yves Hagers, Frank Provenier, Yves Van Belle, Ruben Hamerlijnck and S Serge Barold (2006) PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY. 29(1). p.67-69
Please use this url to cite or link to this publication:
author
organization
year
type
misc (editorialMaterial)
publication status
published
subject
keyword
Lead malfunction, Intracardiac cardioverter-defibrillator (ICD), Defibrillation testing, Noise oversensing, false signals
in
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
PACE-Pacing Clin. Electrophysiol.
volume
29
issue
1
pages
67 - 69
Web of Science type
Editorial Material
Web of Science id
000236179400012
JCR category
ENGINEERING, BIOMEDICAL
JCR impact factor
1.095 (2006)
JCR rank
29/42 (2006)
JCR quartile
3 (2006)
ISSN
0147-8389
DOI
10.1111/j.1540-8159.2006.00290.x
language
English
UGent publication?
no
classification
V
copyright statement
I have transferred the copyright for this publication to the publisher
id
1976336
handle
http://hdl.handle.net/1854/LU-1976336
date created
2012-01-01 12:07:51
date last changed
2016-12-19 15:47:49
@misc{1976336,
  author       = {Stroobandt, Roland and Hagers, Yves and Provenier, Frank and Van Belle, Yves and Hamerlijnck, Ruben and Barold, S Serge},
  issn         = {0147-8389},
  keyword      = {Lead malfunction,Intracardiac cardioverter-defibrillator (ICD),Defibrillation testing,Noise oversensing,false signals},
  language     = {eng},
  number       = {1},
  pages        = {67--69},
  series       = {PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY},
  title        = {Silent lead malfunction detected only during defibrillator replacement},
  url          = {http://dx.doi.org/10.1111/j.1540-8159.2006.00290.x},
  volume       = {29},
  year         = {2006},
}

Chicago
Stroobandt, Roland, Yves Hagers, Frank Provenier, Yves Van Belle, Ruben Hamerlijnck, and S Serge Barold. 2006. “Silent Lead Malfunction Detected Only During Defibrillator Replacement.” Pace-pacing and Clinical Electrophysiology.
APA
Stroobandt, R., Hagers, Y., Provenier, F., Van Belle, Y., Hamerlijnck, R., & Barold, S. S. (2006). Silent lead malfunction detected only during defibrillator replacement. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY.
Vancouver
1.
Stroobandt R, Hagers Y, Provenier F, Van Belle Y, Hamerlijnck R, Barold SS. Silent lead malfunction detected only during defibrillator replacement. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY. 2006. p. 67–9.
MLA
Stroobandt, Roland, Yves Hagers, Frank Provenier, et al. “Silent Lead Malfunction Detected Only During Defibrillator Replacement.” PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY 2006 : 67–69. Print.