Semi-automated micronucleus scoring in cytokinesis-blocked lymphocytes after irradiation

Hubert Thierens UGent, Anne Vral UGent, F De Scheerder, Leo De Ridder and A Tates (1997) INTERNATIONAL JOURNAL OF RADIATION BIOLOGY. 72(3). p.319-324
abstract
A semi-automated scoring procedure for the cytokinesis-block micronucleus assay based on computerized image analysis was validated by comparing results of the automated analysis with those from manual analysis by three experienced scorers of micronuclei. Each object selected by the computer system was also relocated and visually inspected. The validation was based on slides of blood samples after in vitro irradiation with X-rays in the low-dose range up to 2 Gy. Up to 1 Gy a very satisfactory agreement between manual and semi-automated scoring was obtained. At 2 Gy the micronucleus yields obtained by the semi-automated scoring were systematically below the manual data due to a small percentage of false-negatives (3%). Using the appropriate dose-response relationship, the validation study indicates that the results obtained by computerized image analysis are comparable to manual scoring with respect to accuracy. The computerized assay is superior regarding reproducibility. At present, semi-automated scoring of micronuclei is practically feasible for a limited number of slides. For large-scare applications of the semi-automated micronucleus assay in the field of radiation protection an increase of the speed of the system is necessary.
Please use this url to cite or link to this publication:
author
organization
year
type
journalArticle (original)
publication status
published
subject
keyword
DETAILED DESCRIPTION, RADIATION-INDUCED MICRONUCLEI, IMAGE-ANALYSIS, ASSAY
journal title
INTERNATIONAL JOURNAL OF RADIATION BIOLOGY
Int. J. Radiat. Biol.
volume
72
issue
3
pages
319 - 324
Web of Science type
Article
ISSN
0955-3002
DOI
10.1080/095530097143310
language
English
UGent publication?
yes
classification
A1
id
184974
handle
http://hdl.handle.net/1854/LU-184974
date created
2004-01-14 13:41:00
date last changed
2017-03-09 10:30:28
@article{184974,
  abstract     = {A semi-automated scoring procedure for the cytokinesis-block micronucleus assay based on computerized image analysis was validated by comparing results of the automated analysis with those from manual analysis by three experienced scorers of micronuclei. Each object selected by the computer system was also relocated and visually inspected. The validation was based on slides of blood samples after in vitro irradiation with X-rays in the low-dose range up to 2 Gy. Up to 1 Gy a very satisfactory agreement between manual and semi-automated scoring was obtained. At 2 Gy the micronucleus yields obtained by the semi-automated scoring were systematically below the manual data due to a small percentage of false-negatives (3\%). Using the appropriate dose-response relationship, the validation study indicates that the results obtained by computerized image analysis are comparable to manual scoring with respect to accuracy. The computerized assay is superior regarding reproducibility. At present, semi-automated scoring of micronuclei is practically feasible for a limited number of slides. For large-scare applications of the semi-automated micronucleus assay in the field of radiation protection an increase of the speed of the system is necessary.},
  author       = {Thierens, Hubert and Vral, Anne and De Scheerder, F and De Ridder, Leo and Tates, A},
  issn         = {0955-3002},
  journal      = {INTERNATIONAL JOURNAL OF RADIATION BIOLOGY},
  keyword      = {DETAILED DESCRIPTION,RADIATION-INDUCED MICRONUCLEI,IMAGE-ANALYSIS,ASSAY},
  language     = {eng},
  number       = {3},
  pages        = {319--324},
  title        = {Semi-automated micronucleus scoring in cytokinesis-blocked lymphocytes after irradiation},
  url          = {http://dx.doi.org/10.1080/095530097143310},
  volume       = {72},
  year         = {1997},
}

Chicago
Thierens, Hubert, Anne Vral, F De Scheerder, Leo De Ridder, and A Tates. 1997. “Semi-automated Micronucleus Scoring in Cytokinesis-blocked Lymphocytes After Irradiation.” International Journal of Radiation Biology 72 (3): 319–324.
APA
Thierens, H., Vral, A., De Scheerder, F., De Ridder, L., & Tates, A. (1997). Semi-automated micronucleus scoring in cytokinesis-blocked lymphocytes after irradiation. INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 72(3), 319–324.
Vancouver
1.
Thierens H, Vral A, De Scheerder F, De Ridder L, Tates A. Semi-automated micronucleus scoring in cytokinesis-blocked lymphocytes after irradiation. INTERNATIONAL JOURNAL OF RADIATION BIOLOGY. 1997;72(3):319–24.
MLA
Thierens, Hubert, Anne Vral, F De Scheerder, et al. “Semi-automated Micronucleus Scoring in Cytokinesis-blocked Lymphocytes After Irradiation.” INTERNATIONAL JOURNAL OF RADIATION BIOLOGY 72.3 (1997): 319–324. Print.