Femtosecond laser ablation-ICP-mass spectrometry analysis of a heavy metallic matrix: determination of platinum group metals and gold in lead fire-assay buttons as a case study

Frank Vanhaecke UGent, Martín Resano, Joachim Koch, Keith McIntosh and Detlev Günther (2010) JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY. 25(8). p.1259-1267
abstract
Owing to the shorter time interval during which energy is delivered to the sample material, femtosecond (fs) laser ablation is preferable over nanosecond laser ablation for metallic samples. In this project, the influence of various laser parameters-beam diameter, repetition rate and laser fluence-on the ablation of Pb as a heavy metallic matrix using an infrared (lambda = 795 nm) fs-LA system (150 fs pulse duration) was studied. The merits of Ar and He as carrier gases were compared and as He did not provide a substantial improvement in the limits of detection, while deposition of sample material on the window of the ablation chamber was more pronounced, Ar was selected for all further measurements. The effect on the ICP caused by the introduction of various amounts of sample aerosol was studied by monitoring the signal intensity for Ar-38(+). It was shown that maximizing the amount of sample ablated and thus, the amount of sample aerosol introduced into the ICP, did not result in maximum sensitivity, which was rather obtained under 'compromise' conditions. Subsequently, femtosecond LA-quadrupole-based ICP-mass spectrometry (ICP-MS) was used for the determination of traces of the platinum group metals (PGMs) Rh, Pd, Ru, Ir and Pt and of Au in Pb buttons obtained by fire assay of platiniferous ore reference materials. The signal of Pb-204(+) was used as an internal reference, correcting for variations in the laser ablation and transport efficiencies and in the instrument's sensitivity. The spectral interferences established for some of the target nuclides due to the occurrence of Pb2+ ions were successfully overcome by pressurizing the reaction cell with NH3. Quantification versus a calibration curve constructed on the basis of the results obtained for matrix-matched standards (>99% Pb) provided excellent accuracy, superior to those obtained using nanosecond LA-ICP-MS. Also the limits of detection were improved by a factor ranging between 3 and 10 and are <0.010 mu g g(-1) for the most important PGMs (Rh, Pd, Pt) and Au. Several measures, such as the use of a large ablation cell and housing up to 10 Pb buttons, were taken to increase the sample throughput. In the same context, day-to-day reproducibility of the calibration curve was also examined. When recording a 'fresh' calibration curve every day, the average bias between the experimental results and the corresponding reference values was established to be <2.5% for every target element. When using one calibration curve during three consecutive days, the bias still remains <10%, while the sample throughput is increased and analysis of several tens of buttons per day is feasible (10-15 min total analysis time per sample).
Please use this url to cite or link to this publication:
author
organization
year
type
journalArticle (original)
publication status
published
subject
keyword
FRACTIONATION, SIGNAL, NANOSECOND, MS, SILICATE GLASS, PLATINIFEROUS ORES, BRASS, ZONE
journal title
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
J. Anal. At. Spectrom.
volume
25
issue
8
pages
1259 - 1267
Web of Science type
Article
Web of Science id
000280182900005
JCR category
CHEMISTRY, ANALYTICAL
JCR impact factor
4.372 (2010)
JCR rank
5/71 (2010)
JCR quartile
1 (2010)
ISSN
0267-9477
DOI
10.1039/c002746d
language
English
UGent publication?
yes
classification
A1
copyright statement
I have transferred the copyright for this publication to the publisher
id
1113625
handle
http://hdl.handle.net/1854/LU-1113625
date created
2011-02-01 13:39:57
date last changed
2011-03-12 00:30:14
@article{1113625,
  abstract     = {Owing to the shorter time interval during which energy is delivered to the sample material, femtosecond (fs) laser ablation is preferable over nanosecond laser ablation for metallic samples. In this project, the influence of various laser parameters-beam diameter, repetition rate and laser fluence-on the ablation of Pb as a heavy metallic matrix using an infrared (lambda = 795 nm) fs-LA system (150 fs pulse duration) was studied. The merits of Ar and He as carrier gases were compared and as He did not provide a substantial improvement in the limits of detection, while deposition of sample material on the window of the ablation chamber was more pronounced, Ar was selected for all further measurements. The effect on the ICP caused by the introduction of various amounts of sample aerosol was studied by monitoring the signal intensity for Ar-38(+). It was shown that maximizing the amount of sample ablated and thus, the amount of sample aerosol introduced into the ICP, did not result in maximum sensitivity, which was rather obtained under 'compromise' conditions. Subsequently, femtosecond LA-quadrupole-based ICP-mass spectrometry (ICP-MS) was used for the determination of traces of the platinum group metals (PGMs) Rh, Pd, Ru, Ir and Pt and of Au in Pb buttons obtained by fire assay of platiniferous ore reference materials. The signal of Pb-204(+) was used as an internal reference, correcting for variations in the laser ablation and transport efficiencies and in the instrument's sensitivity. The spectral interferences established for some of the target nuclides due to the occurrence of Pb2+ ions were successfully overcome by pressurizing the reaction cell with NH3. Quantification versus a calibration curve constructed on the basis of the results obtained for matrix-matched standards ({\textrangle}99\% Pb) provided excellent accuracy, superior to those obtained using nanosecond LA-ICP-MS. Also the limits of detection were improved by a factor ranging between 3 and 10 and are {\textlangle}0.010 mu g g(-1) for the most important PGMs (Rh, Pd, Pt) and Au. Several measures, such as the use of a large ablation cell and housing up to 10 Pb buttons, were taken to increase the sample throughput. In the same context, day-to-day reproducibility of the calibration curve was also examined. When recording a 'fresh' calibration curve every day, the average bias between the experimental results and the corresponding reference values was established to be {\textlangle}2.5\% for every target element. When using one calibration curve during three consecutive days, the bias still remains {\textlangle}10\%, while the sample throughput is increased and analysis of several tens of buttons per day is feasible (10-15 min total analysis time per sample).},
  author       = {Vanhaecke, Frank and Resano, Mart{\'i}n and Koch, Joachim and McIntosh, Keith and G{\"u}nther, Detlev},
  issn         = {0267-9477},
  journal      = {JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY},
  keyword      = {FRACTIONATION,SIGNAL,NANOSECOND,MS,SILICATE GLASS,PLATINIFEROUS ORES,BRASS,ZONE},
  language     = {eng},
  number       = {8},
  pages        = {1259--1267},
  title        = {Femtosecond laser ablation-ICP-mass spectrometry analysis of a heavy metallic matrix: determination of platinum group metals and gold in lead fire-assay buttons as a case study},
  url          = {http://dx.doi.org/10.1039/c002746d},
  volume       = {25},
  year         = {2010},
}

Chicago
Vanhaecke, Frank, Martín Resano, Joachim Koch, Keith McIntosh, and Detlev Günther. 2010. “Femtosecond Laser ablation-ICP-mass Spectrometry Analysis of a Heavy Metallic Matrix: Determination of Platinum Group Metals and Gold in Lead Fire-assay Buttons as a Case Study.” Journal of Analytical Atomic Spectrometry 25 (8): 1259–1267.
APA
Vanhaecke, Frank, Resano, M., Koch, J., McIntosh, K., & Günther, D. (2010). Femtosecond laser ablation-ICP-mass spectrometry analysis of a heavy metallic matrix: determination of platinum group metals and gold in lead fire-assay buttons as a case study. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 25(8), 1259–1267.
Vancouver
1.
Vanhaecke F, Resano M, Koch J, McIntosh K, Günther D. Femtosecond laser ablation-ICP-mass spectrometry analysis of a heavy metallic matrix: determination of platinum group metals and gold in lead fire-assay buttons as a case study. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY. 2010;25(8):1259–67.
MLA
Vanhaecke, Frank, Martín Resano, Joachim Koch, et al. “Femtosecond Laser ablation-ICP-mass Spectrometry Analysis of a Heavy Metallic Matrix: Determination of Platinum Group Metals and Gold in Lead Fire-assay Buttons as a Case Study.” JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY 25.8 (2010): 1259–1267. Print.