Proof of concept of a new 3D-guided system for a single implant overdenture in the mandible : an in vitro study

Marco Tudts (UGent) , Rani D'haese (UGent) , Geert Hommez (UGent) , Véronique Christiaens (UGent) and Stefan Vandeweghe (UGent)
Author
Organization
Abstract
Purpose: To evaluate the precision and efficiency of a novel guide system for single implant placement in the mandibular symphyses and to evaluate whether the outcome is affected by the level of operator experience. Materials and Methods: A total of 90 implants were placed in three different mandibular cast types (Cawood and Howell class III, IV, and V). For each model, a complete denture was 3D printed. A polyether ether ketone rall with a guide sleeve was embedded in the middle of the denture. To determine the ideal implant position, the sleeve could be moved ina buccolingual direction. Adjustment of implant angulation was possible, and an angle correction of 0, 12, or 24 degrees was available. A total of 30 clinicians were divided into three groups: group 1 (experienced, n = 10), group 2 (beginner, n = 10), and group 3 (inexperienced, n = 10). Each clinician was asked to plan and perform a guided flapless implant placement in the mandibular symphysis. Two preoperative CBCT scans were taken; the first was to verify the planning, and the second was to adjust the planning if needed. Finally, a postoperative CBCT scan was taken to compare the planning to the final implant position. Results: Based on the first CBCT, the clinicians adjusted their planning by an average of 1.66 +/- 1.65 mm coronally, 2.41 +/- 2.44 mm apically, and by a mean angular correction of 6.08 +/- 0.77 degrees. After implant placement, the mean deviation from the planned implant position was 0.87 +/- 0.58 mm at the coronal aspect and 0.98 +/- 0.64 mm at the apical aspect. The mean angular deviation was 6.05 +/- 0.71 degrees. Overall, there were no significant differences in coronal and apical deviation (P > .05) based on the level of experience. In terms of angulation, a significant difference was found in both planning (P= .049) and placement (P=.038) between beginners and experienced clinicians. Conclusions: Guided implant placement of a single implant in the mandibular symphysis using a removable denture with guide sleeve had an acceptable level of accuracy. Clinicians with limited experience spent more time on the procedure, resulting in less angular deviation during implant planning and placement compared to experienced clinicians.
Keywords
dental implants, edentulous mandible, guided surgery, overdenture, single implant, STEREOLITHOGRAPHIC SURGICAL GUIDES, COMPUTED-TOMOGRAPHY, DENTAL IMPLANTS, EDENTULOUS MAXILLA, FOLLOW-UP, ACCURACY, PLACEMENT, SURGERY, TEMPLATE, DENTURE

Downloads

  • (...).pdf
    • full text (Published version)
    • |
    • UGent only
    • |
    • PDF
    • |
    • 642.67 KB

Citation

Please use this url to cite or link to this publication:

MLA
Tudts, Marco, et al. “Proof of Concept of a New 3D-Guided System for a Single Implant Overdenture in the Mandible : An in Vitro Study.” INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, vol. 39, no. 1, 2024, pp. 127–34, doi:10.11607/jomi.10301.
APA
Tudts, M., D’haese, R., Hommez, G., Christiaens, V., & Vandeweghe, S. (2024). Proof of concept of a new 3D-guided system for a single implant overdenture in the mandible : an in vitro study. INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 39(1), 127–134. https://doi.org/10.11607/jomi.10301
Chicago author-date
Tudts, Marco, Rani D’haese, Geert Hommez, Véronique Christiaens, and Stefan Vandeweghe. 2024. “Proof of Concept of a New 3D-Guided System for a Single Implant Overdenture in the Mandible : An in Vitro Study.” INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS 39 (1): 127–34. https://doi.org/10.11607/jomi.10301.
Chicago author-date (all authors)
Tudts, Marco, Rani D’haese, Geert Hommez, Véronique Christiaens, and Stefan Vandeweghe. 2024. “Proof of Concept of a New 3D-Guided System for a Single Implant Overdenture in the Mandible : An in Vitro Study.” INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS 39 (1): 127–134. doi:10.11607/jomi.10301.
Vancouver
1.
Tudts M, D’haese R, Hommez G, Christiaens V, Vandeweghe S. Proof of concept of a new 3D-guided system for a single implant overdenture in the mandible : an in vitro study. INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS. 2024;39(1):127–34.
IEEE
[1]
M. Tudts, R. D’haese, G. Hommez, V. Christiaens, and S. Vandeweghe, “Proof of concept of a new 3D-guided system for a single implant overdenture in the mandible : an in vitro study,” INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, vol. 39, no. 1, pp. 127–134, 2024.
@article{01JHJ2HJDSTH5XE40GRDJPMPZC,
  abstract     = {{Purpose: To evaluate the precision and efficiency of a novel guide system for single implant placement in the mandibular symphyses and to evaluate whether the outcome is affected by the level of operator experience. Materials and Methods: A total of 90 implants were placed in three different mandibular cast types (Cawood and Howell class III, IV, and V). For each model, a complete denture was 3D printed. A polyether ether ketone rall with a guide sleeve was embedded in the middle of the denture. To determine the ideal implant position, the sleeve could be moved ina buccolingual direction. Adjustment of implant angulation was possible, and an angle correction of 0, 12, or 24 degrees was available. A total of 30 clinicians were divided into three groups: group 1 (experienced, n = 10), group 2 (beginner, n = 10), and group 3 (inexperienced, n = 10). Each clinician was asked to plan and perform a guided flapless implant placement in the mandibular symphysis. Two preoperative CBCT scans were taken; the first was to verify the planning, and the second was to adjust the planning if needed. Finally, a postoperative CBCT scan was taken to compare the planning to the final implant position. Results: Based on the first CBCT, the clinicians adjusted their planning by an average of 1.66 +/- 1.65 mm coronally, 2.41 +/- 2.44 mm apically, and by a mean angular correction of 6.08 +/- 0.77 degrees. After implant placement, the mean deviation from the planned implant position was 0.87 +/- 0.58 mm at the coronal aspect and 0.98 +/- 0.64 mm at the apical aspect. The mean angular deviation was 6.05 +/- 0.71 degrees. Overall, there were no significant differences in coronal and apical deviation (P > .05) based on the level of experience. In terms of angulation, a significant difference was found in both planning (P= .049) and placement (P=.038) between beginners and experienced clinicians. Conclusions: Guided implant placement of a single implant in the mandibular symphysis using a removable denture with guide sleeve had an acceptable level of accuracy. Clinicians with limited experience spent more time on the procedure, resulting in less angular deviation during implant planning and placement compared to experienced clinicians.}},
  author       = {{Tudts, Marco and D'haese, Rani and Hommez, Geert and Christiaens, Véronique and Vandeweghe, Stefan}},
  issn         = {{0882-2786}},
  journal      = {{INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS}},
  keywords     = {{dental implants,edentulous mandible,guided surgery,overdenture,single implant,STEREOLITHOGRAPHIC SURGICAL GUIDES,COMPUTED-TOMOGRAPHY,DENTAL IMPLANTS,EDENTULOUS MAXILLA,FOLLOW-UP,ACCURACY,PLACEMENT,SURGERY,TEMPLATE,DENTURE}},
  language     = {{eng}},
  number       = {{1}},
  pages        = {{127--134}},
  title        = {{Proof of concept of a new 3D-guided system for a single implant overdenture in the mandible : an in vitro study}},
  url          = {{http://doi.org/10.11607/jomi.10301}},
  volume       = {{39}},
  year         = {{2024}},
}

Altmetric
View in Altmetric
Web of Science
Times cited: