Types of EMG textile electrodes : a comparative study using PCA
- Author
- Bulcha Belay Etana (UGent) , Benny Malengier (UGent) , Janarthanan Krishnamoorthy and Lieva Van Langenhove (UGent)
- Organization
- Abstract
- Identifying a suitable textile electrode that would be durable and assist in recording high bio-signal quality is crucial in the production of medical devices. Therefore, this study aimed to compare silver-plated-polyamide-embroidered cotton (SPEC), copper-nickel-plated polyester (CNP), and stainless-steel-fabric (SSF) dry textile electromyography (EMG) electrodes through principal component analysis (PCA). Each electrode was also compared against standard silver/silver chloride (Ag/AgCl) gel electrodes. The output was analyzed based on multi variables extracted from EMG features such as root mean square (RMS) voltage, average rectified value (ARV) voltage, kurtosis, and skewness using activation of the bicep, and tibialis anterior muscles group. The SSF electrode outperformed CNP and SPEC electrodes. Without applying any boundary conditions, each textile electrode has signal-to-noise ratio (SNR) values comparable to the standard electrode. The SNR values were 24.38 dB, 17.72 dB, 15.55 dB, and 13.30 dB, for Ag/AgCl, SSF, CNP and SPEC electrodes, respectively. The performance of all the conductive textile electrodes was comparable to that of Ag/AgCl. However, the gel electrode requires skin preparation and has short-term stability, whereas textile electrode materials last longer and can be used for biological signal monitoring at home without the assistance of medical professionals.
- Keywords
- stainless steel, copper-nickel-plated polyester, embroidery, EMG Signal Analysis, PCA
Downloads
-
manuscript (1).pdf
- full text (Published version)
- |
- open access
- |
- |
- 2.19 MB
Citation
Please use this url to cite or link to this publication: http://hdl.handle.net/1854/LU-01GZZYNDBS3D5D19AZ1T84F0Z1
- MLA
- Etana, Bulcha Belay, et al. “Types of EMG Textile Electrodes : A Comparative Study Using PCA.” ENGINEERING PROCEEDINGS (BASEL), vol. 35, no. 1, 2023, doi:10.3390/IECB2023-14560.
- APA
- Etana, B. B., Malengier, B., Krishnamoorthy, J., & Van Langenhove, L. (2023). Types of EMG textile electrodes : a comparative study using PCA. ENGINEERING PROCEEDINGS (BASEL), 35(1). https://doi.org/10.3390/IECB2023-14560
- Chicago author-date
- Etana, Bulcha Belay, Benny Malengier, Janarthanan Krishnamoorthy, and Lieva Van Langenhove. 2023. “Types of EMG Textile Electrodes : A Comparative Study Using PCA.” In ENGINEERING PROCEEDINGS (BASEL). Vol. 35. https://doi.org/10.3390/IECB2023-14560.
- Chicago author-date (all authors)
- Etana, Bulcha Belay, Benny Malengier, Janarthanan Krishnamoorthy, and Lieva Van Langenhove. 2023. “Types of EMG Textile Electrodes : A Comparative Study Using PCA.” In ENGINEERING PROCEEDINGS (BASEL). Vol. 35. doi:10.3390/IECB2023-14560.
- Vancouver
- 1.Etana BB, Malengier B, Krishnamoorthy J, Van Langenhove L. Types of EMG textile electrodes : a comparative study using PCA. In: ENGINEERING PROCEEDINGS (BASEL). 2023.
- IEEE
- [1]B. B. Etana, B. Malengier, J. Krishnamoorthy, and L. Van Langenhove, “Types of EMG textile electrodes : a comparative study using PCA,” in ENGINEERING PROCEEDINGS (BASEL), Online, 2023, vol. 35, no. 1.
@inproceedings{01GZZYNDBS3D5D19AZ1T84F0Z1,
abstract = {{Identifying a suitable textile electrode that would be durable and assist in recording high bio-signal quality is crucial in the production of medical devices. Therefore, this study aimed to compare silver-plated-polyamide-embroidered cotton (SPEC), copper-nickel-plated polyester (CNP), and stainless-steel-fabric (SSF) dry textile electromyography (EMG) electrodes through principal component analysis (PCA). Each electrode was also compared against standard silver/silver chloride (Ag/AgCl) gel electrodes. The output was analyzed based on multi variables extracted from EMG features such as root mean square (RMS) voltage, average rectified value (ARV) voltage, kurtosis, and skewness using activation of the bicep, and tibialis anterior muscles group. The SSF electrode outperformed CNP and SPEC electrodes. Without applying any boundary conditions, each textile electrode has signal-to-noise ratio (SNR) values comparable to the standard electrode. The SNR values were 24.38 dB, 17.72 dB, 15.55 dB, and 13.30 dB, for Ag/AgCl, SSF, CNP and SPEC electrodes, respectively. The performance of all the conductive textile electrodes was comparable to that of Ag/AgCl. However, the gel electrode requires skin preparation and has short-term stability, whereas textile electrode materials last longer and can be used for biological signal monitoring at home without the assistance of medical professionals.}},
articleno = {{6}},
author = {{Etana, Bulcha Belay and Malengier, Benny and Krishnamoorthy, Janarthanan and Van Langenhove, Lieva}},
booktitle = {{ENGINEERING PROCEEDINGS (BASEL)}},
issn = {{2673-4591}},
keywords = {{stainless steel,copper-nickel-plated polyester,embroidery,EMG Signal Analysis,PCA}},
language = {{eng}},
location = {{Online}},
number = {{1}},
pages = {{12}},
title = {{Types of EMG textile electrodes : a comparative study using PCA}},
url = {{http://doi.org/10.3390/IECB2023-14560}},
volume = {{35}},
year = {{2023}},
}
- Altmetric
- View in Altmetric